Filed under: Applications,Help Guides — Tags: , — ups_guru @ 12:35 pm
Current leaks to earth through filters and stray capcitance

Current leaks to earth through filters and stray capcitance

When used in a medical application within the vicinity of a patient, Uninterruptible Power Supplies need to conform to EN60601-1. For details about this please see the Medical UPS or Medical Grade UPS post.

We have a range of Medical Grade UPS up to 1.44KVA. This is big enough for most applications, however what if you require 2KVA or 3KVA? Well the answer is to use a medical grade power conditioner.

These work as they contain an isolation transformer, and rebond the output neutral to the earth conductor (which otherwise passes straight through) thereby eliminating comon mode noise (the raison d’être of a power conditioner) but also ensuring that any earth leakage flows in the secondary of the transformer and not back to the source.

You connect the power conditioner to the wall outlet, the UPS to the power conditioner, and the load to the UPS.

With a medical conditioner, the leakage current is maintained in the secondary

With a medical conditioner, the leakage current is maintained in the secondary

Share this Article:
  • Digg
  • Sphinn
  • del.icio.us
  • Facebook
  • Mixx
  • Google Bookmarks
  • RSS
  • Technorati
  • Blogosphere News
  • LinkedIn
  • MySpace
  • Reddit
  • StumbleUpon
  • laaik.it
  • Live
  • Twitter
  • Twitthis
  • Yahoo! Bookmarks
  • Yahoo! Buzz

No Comments »

No comments yet.

RSS feed for comments on this post. TrackBack URL

Leave a comment

You must be logged in to post a comment.

Powered by WordPress